compact instruments including, ORP and DO DIN 48X48mm, compact body Simple keypad operation including easy calibration procedure Self diagnostics can detect cracks in glass electrode Displays parameters in a sub-menu. Multi operating voltage 100~230V AC Low power consumption, 5VA Memory backup Transmission output adjustable for required pH range CE Marked and in full compliance with relevant EU directives |
||||||
|
|
||||||
| Detailed Specifcations | ||||||
| Product Name: | pH Transmitter. | |||||
| Model: | YPMS-48PH. | |||||
| Measurement Range: | pH; 0~14pH. (Resolution; 0.01pH) mV;-600~600mV. (Resolution; 1mV) Temp;-5-0~100°C. (Resolution; 0.1°C). | |||||
| Repeatability: | ±0.02pH at electric potential input equivalent to pH electromotive force. | |||||
| Transmission Output: | 4~20mADC adjustable setting within 0~14pH for the range specified more than 2pH band with increase by 1pH. | |||||
| Control Output: | Independent output contacts for lower and upper limits, available to switch over to output of maintenance signal or warning signal by key setting. Rating; 250VAC 1A. | |||||
| Calibration: | Simple keypad operation for calibration with two types of buffer solution registered as memory out of pH 1.68, 4.01, 6.86, 9.18, 10.02. | |||||
| Output During Maintenance: | Specify one among Dummy, Hold or Through. | |||||
| Zero Shift: | Adjustable between -1.00 PH~ +1.00pH. | |||||
| Slope Adjustment: | 47.3~63mV/pH. Zero Shift: ±1.00pH. | |||||
| Temp. Calibration Range: | Measured value ±5°C . | |||||
| Temp. Compensation of pH Value: | ±0.100pH/°C . | |||||
| Manual Temp. Compensation: | 0~100°C. | |||||
| Burnout Function: | Active in emergency cases. | |||||
| Power Source: | 100~230V AC, 50/60Hz , 5VA. | |||||
| Mode Control Input: | Switches measure / maintenance mode. | |||||
| Installation: | Panel cut-out; 45 x 45(mm). | |||||
| Panel Thickness: | Max. 4.0mm. | |||||
|
|
||||||