Organic
Pollution Monitor UV Meter
Model OPM-410a
|
|
The
Model OPM-410 is an automated
organic pollution monitor and
performs measurements based on UV
absorptiometry. The instrument is
the result of over 10 years
continuous product development and
this accumulated technology has
resulted in advanced features such
as an improved direct submersion
type detector which eliminates the
need for any extractive sampling
equipment and a construction
designed for low maintenance.
|
 |
|
Features
|
-
Direct
submersion detector
eliminates problems
associated with sample
extraction and pumping.
-
Self-diagnostic
functions continuously
monitor analyser operation.
-
Compact
and lightweight
construction. Detector
weight is only 25% of
previous models. This
simplifies maintenance and
allows the use of corrosion
resistant reinforced
plastic.
-
Accurate
calibration using stability
check function and automatic
temperature compensation.
-
Simplified
span check using built-in
optical filter.
-
Wide
measurement range with
auto-range switching.
-
Output
can be converted to COD from
UV-VIS value.
-
Sample
extraction style analysis
can be performed by
attaching a water tank
receiver.
|
| |
Standard
Specifications
|
Product Name:
|
Organic Pollution
Monitor UV meter. |
| Model: |
OPM-410A. |
| Measurement: |
Organic pollution in effluent. |
| Measurement
Method: |
Dual wavelength absorptiometry (UV rays 254nm Visible rays 365~435nm Continuous measurement).
|
| Measurement
Cell: |
Immersed parallel cell (6mm, 10mm, 25mm). |
| Cell cleaning
Method: |
Automatic surface cleaning of the cell window using a wiper system.
|
| Measurement
Range: |
Absorbance 0~0.5/0~1.0/0~2.0/0~2.5) (auto or manual switching). |
| Ambient
Temperature: |
-5~40°C. |
| Sample
Temperature: |
0~45°C (no freezing). |
| Calibration
Standards |
Zero
Solution:
|
Zero calibration liquid such as distilled water. |
| Span
Solution: |
Potassium hydrogen phthalate. |
| Output
Signals: |
Isolated from ground
UV, UV-VIS, VIS or converted COD value
DC 4~20mA (load 600W or less)
DC 0~1v (load greater than 600W) (available as an option).
|
| Other
Signals |
| Output
Signals: |
Maintenance signal
Power supply cut (short circuit) signal
Low sample water signal
Light source lamp failure signal
In cleaning mode signal
Max measurement value alarm signal
Range switching signal
Analyser fault alarm
(lamp decay, dirty cell window, water leakage, detector decay)
Contact rating for all above: DC 30V 3A, volt-free.
|
| Input
Signals: |
Remote washer start signal
Pulse contact point input to initialize
(Volt free, make time 0.1~1 sec).
|
| Contact
Rating: |
DC 30V 0.3A. |
| Power Supply: |
AC 100V ±10% 50/60Hz. |
| Power
Consumption: |
Approx. 70VA. |
| Measurement
Cell: |
Quartz glass. |
| Drive: |
SUS 316, Urethane rubber. |
| Optional
Body Material : |
AD12 (corrosion resistant aluminum). |
|
| |
Construction
|
Available options:
|
Outdoor installation
(JIS Class 3 Ingress Protection) 192(W) x 120(D) x 491(H) mm.
Self-supporting mounting frame (option): 380(W) x 500(D) x 1500(H) mm
Detector (Submersion type):435(H) x 170 (Æ) mm.
|
| Weight: |
Detector Approx. 6.8kg (not including cable weight). |
| Optional Transmitter Stand: |
Approx. 7kg (support stand
(FRP) 9kg). |
|
| |
Installation
|
| Detector: |
Exclusive fitting (winch mechanism). |
| Transmitter: |
Self supporting stand (standard) or pole mounting. |
| Paint
Finish: |
Munsell 5PB 8/1. |
| Wiring Connection: |
Waterproof plug socket for 15Æ cable x4. |
| Connecting
Cable: |
For between detector and transmitter x1
[15Æ, 10m long (standard), covered with flexible PVC]. |
| Installation: |
Install away from vibration, shocks and corrosive vapors. |
| Air
Connection: |
Air purge facilities are available for the body of the transmitter.
RC ¼, connection.
Air pressure: 0.02 Mpa (0.2
kgf/cm²). |
|
| |
Performance
|
| Repeatability: |
within ±2% FS. |
| Baseline
Drift: |
within ±2% FS. |
| Zero
Drift: |
within ±2%
FS/week. |
| Span
Drift: |
within ±2% FS/week. |
| Response
Speed: |
90% response 30 seconds or less. |
|
| |
Options
|
(1)Components
for modifications to 10mm cell.
(2)Components for modifications to
10mm cell.
(3)Components for sample extraction
style application. |
| |
Related
Equipment
|
| Load
calculation: Use Model CALD-131 for
load calculation. |
| |
Principle
Of Operation
|
|
The
instrument uses a low pressure
mercury lamp powered by a high
stability circuit. Light from this
source passes through the cell and
then through a UV-pass filter (l =
254nm) and VIS-pass filter (l =
365~435 nm).
The analyser calculates the
difference between AUV and AVIS and
generates an output signal via a V/I
converter. By using this system,
obscuration due to the presence of
mud in the sample can be compensated
for and the instrument can
effectively determine the
concentration of organic pollutants
in the sample.
|
| |
Notes
|
-
Unit can operate on both 50 or 60 Hz
power.
-
Unit can operate on
non-standard power supplies by using
a step down transformer available
separately (such as Model ZP type).
-
Three types of transmission
outputs are available. These are UV,
UV-VIS and VIS or converted COD
value (VIS and converted COD value
can be changed from the key-pad).
-
AUTO is the automatic range
switch of absorptiometer
0.5/1.0/2.0/2.5 and is available as
standard equipment. Use by fixed
range when combine with total load.
-
Maximum length for
transmitter-sensor cable is 30m.
-
If the FRP support stand is
not required, indicate
“non-support stand” under
installation style.
-
For sample
extraction/collection style
installations, it is unnecessary to
arrange the winch equipment and also
the UV meter sampling water tank,
Model OPM-410W.
-
For easier installation
method, make the best use of OPM-410
type’s advantages, we suggest the
hanging style configuration with two
of 3.5mm chains (if this
specification is selected, the
detector will be provided with two
eye bolts).
-
In
case of no wall around the guide pipe of the winch
equipment, the detector may rotate. In these situations,
select ‘dual guide pipes style’. The same is also
applicable to fast moving stream applications.
|
|

|